DLRO10X Product Page
|
|
Menu system interface
Alpha-numeric keypad
Multiple operating modes
User settable pass limits
Protected to 600 V
Printer output & memory
Real-time result download
DLRO10 and DLRO10X set the standards for low resistance measurement. DLRO10 and DLRO10X are fully automatic instruments,
selecting the most suitable test current up to 10A d.c. to measure resistance from 0.1µΩ to 2000Ω, on one of seven ranges.
For users who desire more control over the measurement process, DLRO10X uses a menu system controlled by a two-axis paddle to allow the user to
manually select the maximum test current. DLRO10X also adds real time download of results and on board storage for later download to a PC.
Both instruments are built into a strong, lightweight case that is equally at home in the field or in the laboratory. Light enough to be worn around the neck,
they are small enough to be taken into areas that were previously too small to access. The DLRO10X has a large, backlit LCD display. Normally, measurements are made with forward
and reverse currents to cancel the effects of any standing voltages across the test sample. The average value is then displayed within 3 seconds, to a basic accuracy of 0.2%.
DLRO10X displays both forward and reverse measurements as well as the average of the two. DLRO10X allows the user to set high and low pass limits,
thereby enabling simple go-no-go testing. At the end of a test DLRO10X will store the test results, as well as any notes relevant to the test.
To assist operator safety and ease of use, both instruments are supplied complete with a pair of duplex handspikes with 1.2 m (4 ft) leads. One of the probes is fitted with LED's,
which duplicate indicators on the instrument display indicating that all four contacts have been made, the presence of a high voltage across the load, and the presence of current
flow while a load is discharging. A full range of test leads is available with probes, clamps and Kelvin clips. Nickel Metal Hydride (NiMH) batteries power the instruments.
The battery packs are interchange-able so that an exhausted battery may be recharged using the external charger supplied while testing continues using a spare pack. Although
full charging will take 4 hours, a fast charge mode allows the battery to be 90% charged within 2½ hours from a 12 volt battery or from a standard 120/230 V AC supply via
the supplied charger. The battery pack contains its own battery state indicator, which allows the charge-state to be monitored, even without being connected to the instrument.
DLRO10X is fitted with RS232 communications that will allow results to be downloaded in real time or stored for later retrieval. Up to 700 sets of results may be stored within
DLRO10X complete with notes containing up to 200 characters which may be added using the on board keypad. These results can also be downloaded to a PC.
Measurement:
Mode: Auto, Continuous, High Power, Unidirectional
Control: Fully Automatic/Manual
Speed: <3s for forward & reverse current and to display average
Display:
Measurement: 41/2 digit seven segment LED
Range & Safety: Large backlit LCD
Test Method:
Single cycle reversing d.c. ratiometric measurement -average result display
Test Current:
Accuracy: ±10%
Stability: <10 ppm per second
Maximum Lead Resistance:
100 mΩ total for 10A operation irrespective of battery condition.
Voltmeter input impedance:
> 200 kΩ
Hum Rejection:
Less than 1% ±20 digits additional error with 100 mV peak 50/60 Hz. on the potential leads. Warning will show if hum or noise exceeds this level.
Data:
Transfer: Real Time or from storage via RS232
Storage: 700 Tests
Memo Field:Up to 200 characters per test via integral alphanumeric keypad
Battery:
Capacity: 7 Ah NiMH rechargeable
Life:Typically 1000 x 10 A tests before recharge
Recharge Via External 90V - 260V 50/60 Hz charger or from 12 to 15V d.c. supply
Charging Rate:
Standard: 2.5 hours to 90% capacity, 4 hrs for full charge
Temperature:
Operation: +5ºC to +45ºC (41ºF to 113ºF) at full specification -10°C to +50°C (14°F to 122°F) at reduced accuracy
Storage:
-30ºC to +70ºC (-22ºF to 158ºF)
Coefficient: <0.01% per ºC over range 5ºC to 40ºC (<0.006% per ºF from 41ºF to 104ºF)
Slow Charging: +10°C to +45°C (50°F to 113°F)
Humidity (max):
90% RH @ 40ºC (104ºF) non-condensing
Altitude (max):
2000m (6562 ft) to full safety specifications
Safety:
In accordance with IEC61010-1 600V Category III - only when DH6 leads are used
EMC:
In accordance with IEC61326-1
Dimensions:
220 x 100 x 237 mm (8.6 x 4 x 9.5 in)
Weight:
2.6 kg (5 3/4 lb.) including battery module.
Included Accessories
|
7 Ah NiMH battery module.
|
6340-101
|
|
|
|
DH4 Duplex handspikes (2).
One with indicator lights. 1.2m / 4 ft
|
6111-503
|
|
|
|
Battery charger for operation from 115/230 V.
50/60Hz supply.
|
6280-333
|
|
|
|
Cigar lighter adapter for battery charging.
|
6280-332
|
|
|
|
User guide.
|
6172-473
|
|
|
Optional Accessories
|
|
|
Carrying case for DLRO10X and all standard accessories.
|
6380-138
|
|
|
|
Carrying case for optional lead sets.
|
18313
|
|
|
|
Calibration Shunt,10Ω, current rating 1 mA.
|
249001
|
|
|
|
Calibration Shunt, 1Ω, current rating 10 mA.
|
249002
|
|
|
|
Calibration Shunt, 100 mΩ current rating 1A.
|
249003
|
|
|
|
Certificate of Calibration for Shunts, NIST.
|
CERT-NIST
|
|
|
|
Replacement tips for DH4, DH5 and DH6 handspikes.
Needle point
|
25940-012
|
|
|
|
Replacement tips for DH4, DH5 and DH6 handspikes.
Serrated end
|
25940-014
|
Optional Test Leads
|
|
|
DH5 straight duplex handspikes (2).
One has indicator lights. 2.5m/8ft
|
6111-517
|
|
|
|
DH6 Duplex handspikes (2).
suitable for working on 600 V. systems. 2.5m/8ft
|
6111-518
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 2m/7ft
|
242011-7
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 2.5m/8ft
|
6111-022
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 5.5m/18ft
|
242011-18
|
|
|
|
Duplex Handspike (1) with spring loaded helical contacts.
only one lead supplied. 6m/20ft
|
6111-023
|
|
|
|
Duplex Handspike (1) with spring loaded helical contacts.
only one lead supplied. 9m/30ft
|
242011-30
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 2m/7ft
|
242002-7
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 5.5m/18ft
|
242002-18
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 9m/30ft
|
242002-30
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
2m/7ft
|
242004-7
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
5.5m/18ft
|
242004-18
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
9m/30ft
|
242004-30
|
|
|
|
Duplex handspikes with replaceable needle points
2m/7ft
|
242004-30
|
|
|
|
Duplex 1.27 cm (1/2 ") Kelvin Clips. (2) gold plated
2m/7ft
|
241005-7
|
|
|
|
Duplex 1.27 cm (1/2 ") Kelvin Clips. (2) silver plated
2m/7ft
|
242005-7
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
2m/7ft
|
242006-7
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
5.5m/18ft
|
242006-18
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
9m/30ft
|
242006-30
|
|
|
|
Single handspike (1) for potential measurement.
2m/7ft
|
242021-7
|
|
|
|
Single handspike (1) for potential measurement.
5.5m/18ft
|
242006-18
|
|
|
|
Single handspike (1) for potential measurement.
9m/30ft
|
242021-30
|
|
|
|
Current clip (1) for current connections.
2m/7ft
|
242041-7
|
|
|
|
Current clip (1) for current connections.
5.5m/18ft
|
242041-18
|
|
|
|
Current clip (1) for current connections.
9m/30ft
|
242041-30
|
| Keep up-to-date with the latest documentation for your device. |
| |
| Category |
Title |
Size |
|
| |
| Data Sheet |
The latest information data sheet |
| |
DLRO10X Data Sheet
|
724 KB |
|
| User Guide |
| |
DLRO10X User Guide
|
707 KB |
|
|
|
|
|
INTRODUCING THE DLRO10X
Menu system interface
Alpha-numeric keypad
Multiple operating modes
User settable pass limits
Protected to 600 V
Printer output & memory
Real-time result download
DLRO10 and DLRO10X set the standards for low resistance measurement. DLRO10 and DLRO10X are fully automatic instruments,
selecting the most suitable test current up to 10A d.c. to measure resistance from 0.1µΩ to 2000Ω, on one of seven ranges.
For users who desire more control over the measurement process, DLRO10X uses a menu system controlled by a two-axis paddle to allow the user to
manually select the maximum test current. DLRO10X also adds real time download of results and on board storage for later download to a PC.
Both instruments are built into a strong, lightweight case that is equally at home in the field or in the laboratory. Light enough to be worn around the neck,
they are small enough to be taken into areas that were previously too small to access. The DLRO10X has a large, backlit LCD display. Normally, measurements are made with forward
and reverse currents to cancel the effects of any standing voltages across the test sample. The average value is then displayed within 3 seconds, to a basic accuracy of 0.2%.
DLRO10X displays both forward and reverse measurements as well as the average of the two. DLRO10X allows the user to set high and low pass limits,
thereby enabling simple go-no-go testing. At the end of a test DLRO10X will store the test results, as well as any notes relevant to the test.
To assist operator safety and ease of use, both instruments are supplied complete with a pair of duplex handspikes with 1.2 m (4 ft) leads. One of the probes is fitted with LED's,
which duplicate indicators on the instrument display indicating that all four contacts have been made, the presence of a high voltage across the load, and the presence of current
flow while a load is discharging. A full range of test leads is available with probes, clamps and Kelvin clips. Nickel Metal Hydride (NiMH) batteries power the instruments.
The battery packs are interchange-able so that an exhausted battery may be recharged using the external charger supplied while testing continues using a spare pack. Although
full charging will take 4 hours, a fast charge mode allows the battery to be 90% charged within 2½ hours from a 12 volt battery or from a standard 120/230 V AC supply via
the supplied charger. The battery pack contains its own battery state indicator, which allows the charge-state to be monitored, even without being connected to the instrument.
DLRO10X is fitted with RS232 communications that will allow results to be downloaded in real time or stored for later retrieval. Up to 700 sets of results may be stored within
DLRO10X complete with notes containing up to 200 characters which may be added using the on board keypad. These results can also be downloaded to a PC.
Measurement:
Mode: Auto, Continuous, High Power, Unidirectional
Control: Fully Automatic/Manual
Speed: <3s for forward & reverse current and to display average
Display:
Measurement: 41/2 digit seven segment LED
Range & Safety: Large backlit LCD
Test Method:
Single cycle reversing d.c. ratiometric measurement -average result display
Test Current:
Accuracy: ±10%
Stability: <10 ppm per second
Maximum Lead Resistance:
100 mΩ total for 10A operation irrespective of battery condition.
Voltmeter input impedance:
> 200 kΩ
Hum Rejection:
Less than 1% ±20 digits additional error with 100 mV peak 50/60 Hz. on the potential leads. Warning will show if hum or noise exceeds this level.
Data:
Transfer: Real Time or from storage via RS232
Storage: 700 Tests
Memo Field:Up to 200 characters per test via integral alphanumeric keypad
Battery:
Capacity: 7 Ah NiMH rechargeable
Life:Typically 1000 x 10 A tests before recharge
Recharge Via External 90V - 260V 50/60 Hz charger or from 12 to 15V d.c. supply
Charging Rate:
Standard: 2.5 hours to 90% capacity, 4 hrs for full charge
Temperature:
Operation: +5ºC to +45ºC (41ºF to 113ºF) at full specification -10°C to +50°C (14°F to 122°F) at reduced accuracy
Storage:
-30ºC to +70ºC (-22ºF to 158ºF)
Coefficient: <0.01% per ºC over range 5ºC to 40ºC (<0.006% per ºF from 41ºF to 104ºF)
Slow Charging: +10°C to +45°C (50°F to 113°F)
Humidity (max):
90% RH @ 40ºC (104ºF) non-condensing
Altitude (max):
2000m (6562 ft) to full safety specifications
Safety:
In accordance with IEC61010-1 600V Category III - only when DH6 leads are used
EMC:
In accordance with IEC61326-1
Dimensions:
220 x 100 x 237 mm (8.6 x 4 x 9.5 in)
Weight:
2.6 kg (5 3/4 lb.) including battery module.
Included Accessories
|
7 Ah NiMH battery module.
|
6340-101
|
|
|
|
DH4 Duplex handspikes (2).
One with indicator lights. 1.2m / 4 ft
|
6111-503
|
|
|
|
Battery charger for operation from 115/230 V.
50/60Hz supply.
|
6280-333
|
|
|
|
Cigar lighter adapter for battery charging.
|
6280-332
|
|
|
|
User guide.
|
6172-473
|
|
|
Optional Accessories
|
|
|
Carrying case for DLRO10X and all standard accessories.
|
6380-138
|
|
|
|
Carrying case for optional lead sets.
|
18313
|
|
|
|
Calibration Shunt,10Ω, current rating 1 mA.
|
249001
|
|
|
|
Calibration Shunt, 1Ω, current rating 10 mA.
|
249002
|
|
|
|
Calibration Shunt, 100 mΩ current rating 1A.
|
249003
|
|
|
|
Certificate of Calibration for Shunts, NIST.
|
CERT-NIST
|
|
|
|
Replacement tips for DH4, DH5 and DH6 handspikes.
Needle point
|
25940-012
|
|
|
|
Replacement tips for DH4, DH5 and DH6 handspikes.
Serrated end
|
25940-014
|
Optional Test Leads
|
|
|
DH5 straight duplex handspikes (2).
One has indicator lights. 2.5m/8ft
|
6111-517
|
|
|
|
DH6 Duplex handspikes (2).
suitable for working on 600 V. systems. 2.5m/8ft
|
6111-518
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 2m/7ft
|
242011-7
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 2.5m/8ft
|
6111-022
|
|
|
|
Duplex Handspikes (2) with spring.
loaded helical contacts. 5.5m/18ft
|
242011-18
|
|
|
|
Duplex Handspike (1) with spring loaded helical contacts.
only one lead supplied. 6m/20ft
|
6111-023
|
|
|
|
Duplex Handspike (1) with spring loaded helical contacts.
only one lead supplied. 9m/30ft
|
242011-30
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 2m/7ft
|
242002-7
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 5.5m/18ft
|
242002-18
|
|
|
|
Straight Duplex Handspikes (2)
Heavy Duty with fixed contacts. 9m/30ft
|
242002-30
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
2m/7ft
|
242004-7
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
5.5m/18ft
|
242004-18
|
|
|
|
Duplex Heavy Duty 5cm (2") C-Clamps. (2)
9m/30ft
|
242004-30
|
|
|
|
Duplex handspikes with replaceable needle points
2m/7ft
|
242004-30
|
|
|
|
Duplex 1.27 cm (1/2 ") Kelvin Clips. (2) gold plated
2m/7ft
|
241005-7
|
|
|
|
Duplex 1.27 cm (1/2 ") Kelvin Clips. (2) silver plated
2m/7ft
|
242005-7
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
2m/7ft
|
242006-7
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
5.5m/18ft
|
242006-18
|
|
|
|
Duplex 3.8 cm (11/2") Kelvin Clips. (2)
9m/30ft
|
242006-30
|
|
|
|
Single handspike (1) for potential measurement.
2m/7ft
|
242021-7
|
|
|
|
Single handspike (1) for potential measurement.
5.5m/18ft
|
242006-18
|
|
|
|
Single handspike (1) for potential measurement.
9m/30ft
|
242021-30
|
|
|
|
Current clip (1) for current connections.
2m/7ft
|
242041-7
|
|
|
|
Current clip (1) for current connections.
5.5m/18ft
|
242041-18
|
|
|
|
Current clip (1) for current connections.
9m/30ft
|
242041-30
|
| Keep up-to-date with the latest documentation for your device. |
| |
| Category |
Title |
Size |
|
| |
| Data Sheet |
The latest information data sheet |
| |
DLRO10X Data Sheet
|
724 KB |
|
| User Guide |
| |
DLRO10X User Guide
|
707 KB |
|
|
|
|
|
INTRODUCING THE DLRO10X
|
|
|
 |
|
 |
|
|